Loading documents preview...
Sensors-transducers
Presented by: I&C Technician
Introduction to process measurement • The purpose of measurement • Kinds of signals • Remote vs local display • Error in measurement system • Calibration • Calibration laboratory • Field calibration
Function of sensor 1. An instrument must sense the process to be measured. part of the instrument must stretch bend, or change position or characteristics when affected by process. 2. The instrument must convert this response into an electrical or pneumatic signals.
CONTROL LOOP
transmitter
Function of transducer • The process sensing element is usually a bellow, a diaphragm or a bourden tube. • The primary conversion element converts the mechanical action of a process sensing element to an electrical signal, usually an electrical resistance or a voltage. • The secondary conversion element is usually an electronic circuit that produces a standard signal according to the need of the control system.
Sensors transducers • • • • • • • •
Pressure sensors. Flow sensors. Thermometers. Liquid level sensors. Vibration and speed sensors. Proximity and displacement sensors. Fire, Flame and smoke detectors. Analyzers.
Transducer principle 1. 2. 3. 4. 5. 6. 7. 8.
Capacitive. Inductive. Piezoelectric. Resistive. Potentiometer. Strain gage. Photoconductive. Thermoelectric.
PRESSURE
BOURDEN TUBE PRESSURE SENSOR
BELLOWS DIAPHRAGM STRAIN GAUGE CAPACITVE INDUCTIVE
PRESSURE TRANSDUCER
POTENTIOMETRIC MAGNETIC PIEZOELECTRIC PHOTOELECTRIC
Bourdon tubes
bellows
diaphragm
Strain gauge
Capacitive
potentiometer
piezoelectric
inductive
Working on transmitter • A measurement device is made up of two parts • Primary element • Transmitter • Converts signal (Transducer) • Transmits signal
flow meters • Flow is the motion of fluid. • Flow rate is the time rate of motion fluid. • Total flow is the rate integrated over a specific time interval. • Flow meter is a flow rate transducer. • Laminar flow. • Turbulent flow. • Reynolds no.
Units of measurement • Cubic foot/minuet. • Gallon/minuet. • Liter/sec.
FLOW
DIFFERENCIAL PRESSURE
PRIMARY ELEMENT
MECHANICAL FLOWMETER
POSITIVE DISPLACEMENT
ELECTRONIC FLOWMETER
TURBINE FLOWMETER MEGNATIC FLOWMETER
ROTARY FLOWMETER
VORTEX FLOWMETER
ULTRASONIC FLOWMETER
Common Devices To Measure Flow • Orifice
plate
• Veturi tube • Flow nozzle
Types and use of orifice plate 1. Concentric opening in center. It is for normal use.
2. Eccentric Opening slightly off center. Used for the fluid or gas which has solid particles.
Types and use of orifice plate 3 . Segmental semi circular in shape. It is for used for slurries.
Venturi Tube
Flow Nozzle
Variable area instruments
Rota meter
Positive displacement flow meters • Trap a known volume of fluid and allow it to pass from meter inlet to outlet. • Total flow measurement. • They have accurate,precise,and have a wide flow range. • They are expensive and object to corrosion. • They are used for water meters and gasoline pumps.
PD flow meters types • Piston flow meter • Notating-disc flow meter • Lobed impeller and oval flow meter • Rotary vane meter • Turbine flow meter
Nutating and rotary-vane flow meter
Piston type flow meter
Lobed impeller and oval flow meter
Turbine flow meter
Magnetic flow meter
Vortex flow meter
Ultrasonic flow meter
temperature Bimetallic and liquid filled system
Resistance temperature detector
thermister
thermocouple
Temperature
Hotness or Coldness of a Substance
Temperature Units – Celsius
°C
– Fahrenheit °F – Kelvin
°K
Temperature Measuring Instruments
Instrument Inserted or Attached Located at Distance
Pyromete r
Thermomete rs
Basic Types Of Thermometers
Bimetallic – From –75 °C to 540 °C
Liquid In Glass – From -120 °C to 320 °C
Filled System – From -273 °C to -185 °C – From 650 °C to 760 °C
Metals expand on heating and contract on cooling.
Each metal has a coefficient of thermal expansion
Bimetallic strip bends toward metal of lower rate of thermal expansion, on heating.
Bimetallic Thermometers
– Spiral – Helix, Multiple Helix
Bimetallic Thermometers (Cont.) Office
and Home Use
Bimetallic Thermometers (Cont.) Industrial
Use
Bimetallic Thermometers (Cont.) Advantages : Inexpensive,
Read
Reasonably
Carefully Best
Rugged and Easy to
Accurate, Handle
for Local Use
RTDs Some Metal’s Resistance Increase on Increase of Temperature Metals Used in RTD Must Be
Pure
Uniform Quality
Stable With Given Temperature Rage
Able to Reproduce Resistance Temp. Reading
RTDs
Temperature, °C 0 200 400
Metal
Resistance
Platinum Copper Nickel Platinum Copper Nickel Platinum Copper Nickel
100 100 100 175 186 250 250 275 485
RTDs
RTD act just like a variable resistor Used in resistance bridge circuit Output voltage resistance of RTD Heating or cooling cause bridge unbalance Lead wire error 250 ft copper wire has 1 ohm resistance. If resistance 139 ohm temp. will be 100°C exclude 2 ohm then R=137 which show 94.4 °C error 5.1 %
RTDs
Compensating Lead Error – Use of Third Wire, RTD Wire Should Be Opposite Arm – Error With 2 Lead Wires 4°C, With Three Lead 0.1 °C and by Fourth Wire Error 0.03°C
RTDs
–Construction of RTD
RTDs –Measuring Surface Temperature RTDS –Protecting Well for RTDs
RTDs
Advantages of RTD – Extremely accurate – Measure temperature within 0.15 °C – Fast in response and small in size Disadvantages of RTD – Expansive and complex – Require bridge circuit, power supply, meter
Thermistors
Special Type Of Resistors
High temp. resistance coefficient
Negative as well as positive Temperature Coefficient
Resistance Increases Temp. Decreases
Advantages – Three & four wire bridge circuit not required – Two lead wire provide sufficient accuracy – Less expansive and less complicated Disadvantages – Lack RTD accuracy.
Thermocouple
Two Different Metals or Alloy – Two Junction • Reference Junction ( to be at constant temperature) – Measuring Junction • Heating measuring junction produce voltage across MJ & reference junction
Thermocouple
Extension Wires Thermocouple wires Same as thermocouple ( iron constantan, copper constantan Other material for platinum thermocouple Copper or copper alloy
Thermocouple • Construction of Thermocouple
Thermocouple
Advantages – In Expansive, Rugged, Accurate – Do not use bridge circuit – Wide temperature range –273 to 2800 oC
Disadvantages – Low voltage. – Reference junction kept to be constant
Level
Sight glass Pressure head instruments
displacer Differential pressure
Float switches
Electrical instruments
vibration ultrasonic
Definition of Some Points Datum point:
A fixed reference point from which level is measured
Buoyancy:
Upward force exerted by liquid on an object submerged & floating in it
Buoyant force:
Net upward force exerted by liquid on submerged or floating object equal the weight of the displaced liquid. A nearly stationary object whose purpose is to displace liquid, used for liquid measurement.
Displacer:
Level Measurement by Float
A float, cable, sliding pointer arrangement.
A fixed scale is mounted on the exterior of the tank
A cable connects the float with the pointer through two pullies.
Pointer move up and down and also serve as counter weight.
Float type level indicators are installed on the open tanks and the tank of the atmospheric pressure.
Float-based level switch
Level Measurement by Displacer Displacer do not ride on the surface of the liquid but they are design to submerge as the level rises. Displacer mechanism transfer the slight mechanical motions due to loss of weight which can be used in instruments to indicate level.
Torque Tube Displacers
The most common type of displacer mechanism is a torque tube displacer. Torque tubes are useful when level measurement is needed on the liquid under pressure. As the liquid level changes, the buoyant force moves the displacer slightly and twist the torque tube. The motion of the torque tube makes the pointer move to indicate the liquid level on the scale.
Displacement level
Level Measurement Using Differential Pressure Instruments
Differential pressure measurement is suited to level measurement in closed vessels
When liquid level is measured in an open tank, piping from the lower end of the tank is connected to the high pressure side of the transmitter and the low pressure side is vented.
When liquid level is measured in a closed tank the effect of tank pressure is compensated by connecting a line from the top of the tank to the low pressure side of the differential.
The pressure due to the height of the liquid is a measure of its level.
Hydrostatic tank gauging
Differential pressure cell design
Ultrasonic level
analyzers Hydrazine monitor
Sodium monitor Dissolve oxygen monitor Flue gas oxygen monitor
Silica monitor Conductivity monitor
Ph monitor
hydrazine
Main component
sensor
Sample and solution flow
Calibration sequence
Sodium • Reagent solution • Standard solution • Etch solution
Dissolved Oxygen Sensor • Oxygen molecules diffuse through the membrane based on the partial pressure of oxygen. • ORP Reaction – Oxygen is reduced to hydroxide ions (OH-) at the gold cathode. – Silver is oxidized to silver bromide (M-4) or silver chloride (M-4HD) at the silver anode
• The current flow is directly proportional to the concentration of Dissolved Oxygen in solution
Calibration • Zero calibration • Air calibration
Silica monitoring
Sample and solution flow
Calibration and maintenance • Reagent solution • Standard solution
Flue gas oxygen monitor
installation
Main probe component
Cell and tube assembly
Diffuser and detector
Inner probe assembly
Calibration setup
CONDUCTIVITY ANALYZERS
Measuring principle Conductance is a measure ability of electric circuit to pass current.
G = 1/R(siemense) Na+ Cl-
Variables affecting conductivity • Temperature • Concentration • contamination
Conductivity probes • Electrode type • Inductive type
Na+ Cl-
Typical Conductivity Values 10% NaOH
355,000
Sea Water
50,000
Mains Water Supply
50
Boiler Feed Water
1
Distilled Water
0.5
Ultra Pure Water
0.055
0
1
100
10,000
1,000,000
CONDUCTIVITY IN HUB PLANT
CONDENSER/EXTRACTION PUMPS
STATOR COOLING WATER
BOILER FEED WATER
Calibration Zero standard solution
Span standard solution
Grab sample
Conductivity ranges vs. resistance input
Decade box
PH monitor
pH Theory • pH is a Unit of Measurement – pH = Power of Hydrogen (H+) – Defined as the Negative Logarithm of Hydrogen Ion Activity • pH = log (1/H+)
• Used for Determining the Acidity or Alkalinity of an Aqueous Solution – Practical pH Scale for Industrial Instrumentation • 0 - 14 pH
pH Sensor Components • pH Measuring Electrode pH Measuring Electrode
Reference Cell
– Purpose is to Develop a Millivolt Potential Directly Proportional to the Free Hydrogen Ion Concentration in an Aqueous Solution
• Reference Cell KCl Buffered to 7 pH KCl Gel Ag/AgCl Element
pH Sensitive Glass
Liquid Junction
– Purpose is to Maintain a Constant Reference Potential Regardless of pH Change or Other Ionic Activity in the Solution
• Reference Cell Liquid Junction – Purpose is to Maintain Electrical Contact Between the Reference Electrode and the Measuring Electrode by way of the Solution
pH is a Potentiometric Measurement pH Measuring Electrode
Reference Electrode
• The Measuring System consists of a pH Measuring Electrode and Reference Electrode • The Potential Difference Between the Two Electrodes is a Function of the pH Value of the Measured Solution • The Solution Must Be Conductive and is Part of the Electrical Circuit
Ion Concentration Acid
Neutral
Alkaline
pH 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14
Hydrogen Ion (H+) 1 0.1 0.01 0.001 0.0001 0.00001 0.000001 0.0000001 0.00000001 0.000000001 0.0000000001 0.00000000001 0.000000000001 0.0000000000001 0.00000000000001
Hydroxyl Ion (OH-) 0.00000000000001 0.0000000000001 0.000000000001 0.00000000001 0.0000000001 0.000000001 0.00000001 0.0000001 0.000001 0.00001 0.0001 0.001 0.01 0.1 1
Typical pH Values 4.0% NaOH Crackers Blood Pure Water Milk Beer Tomatoes Oranges Lemons 4.9 % H2SO4
0.0
1.0
2.0
3.0
4.0
5.0
6.0
7.0
pH
8.0
9.0 10.0 11.0 12.0 13.0 14.0
Ph measurement in hub plant
Boiler feed water
Waist water treatment
calibration
Standard buffers
Grab sample
Fire and smoke detectors
Ionization sensors
Photo sensors
Infrared and ultraviolet sensors
Thermal sensors
Flam detector heat generated
Ionization
radiation
Fire and smoke detectors
Ionization sensors
Photo sensors
Infrared and ultraviolet sensors
Thermal sensors
Flam detector heat generated
Ionization
radiation
Proximity and displacement sensors Capacitive sensors
Inductive sensors
Magnetic sensors
LVDT sensors
Mechanical limit switches
Optical sensors
Ultrasonic sensors
Digital output and encoders