Theory of Scanning Electron Microscope
Lettuce Field(16M DRAM)
Hitachi HighHigh- Technologies Corporation
Theory of Scanning Electron Microscope Lump (Illumination Source)
Electron Source
Condenser Lens
Condenser Lens
Objective Lens
Objective Lens
Projection Lens
O M
Scanning
SE Detector
Sample Fluorescent screen
Screen Image
Deflection Coils
Sample
Sample
Image
TEM
Image
SEM
Difference among OM, TEM and SEM
C R T
The Example of Observation of Alga (Mesostigma (Mesostigma)) by Optical Microscope and SEM
Accelerating Voltage: 5kV
Optical Microscope
SEM
The Difference in the Depth of Focus of Optical Microscope and SEM Fine Electron Beam with Small Incident Angle
α2
α1 Depth of Focus (Shallow)
Depth of Focus ( Deep Deep) )
Specimen
Optical Microscope ( α 1≒ 1rad 1rad) )
SEM ( α 2= 10 -2∼ 10 -3rad rad) )
Theory of Scanning Electron Microscope High Voltage
Camera Filament
Electron Gun
Wehnelt Anode Condenser Lens Deflection Coils
CRT Mag. Control Deflection Coils Deflection Amplifier Scanning Electron Beam
Objective Lens Specimen Chamber
SE Detector
Image Signal
Amplifier
Specimen
Vacuum Pump
Configuration of a scanning electron microscope
Theory of Scanning Electron Microscope
Scanning Electron Beam of CRT L
Scanning Electron Probe Scanning (X) Scanning(Y)
l S E M
Pixel
Specimen
CRT Magnification :( M)=L / l
Magnifying mechanism in the SEM
What is Magnification on Hitachi’s SEM
Magnification=DD/DS
Scan coil
DD 4 x 5 inch Polaroid unit (120 x 90mm)
Sample DS
Theory of Scanning Electron Microscope Al Coating Layer Primary Electron Beam Phosphors
Scintillator Photomultiplier Photo Multiplier Tube
Secondary Electron
Photons Light Guide
Specimen
Signal
+10kV CRT
Secondary electron detection system
What is resolution on SEM? Example : Digital Camera
Magnified Could not identify the gap between point and point
Magnified
What is resolution on SEM? On SEM : Identify finest gap between 2 particles
Can not identify
Can identify
What is resolution on SEM?
100nm
S-4800
S-5500
Specimen Vacc
: Au particle : 15kV
Specimen Vacc
: Pt particle : 30kV
Mag.
: 220kX
Mag.
: 800kX
Resolution
: 1.0nm
Resolution
: 0.4nm
Theory of Scanning Electron Microscope Characteristic X-Ray Backscattered Electron
↓
Primary Electron Beam
↓ ↓
Secondary Electron
Cathodeluminescence ~10nm
Secondary Electron Detector (Excitation Volume for Secondary Ele ctron Emission)
Electron Beam Induced Current
Specimen Current Transmitted Electron Transmitted (Scattered) Electron
The primary electron beambeam-specimen interaction in the SEM
Theory of Scanning Electron Microscope
Backscattered Electron −
−
Secondary Electron −
Vaccum −
Sample(Metal)
Simons,et.al
Theory of Scanning Electron Microscope
Quantity of Electrons
Secondary Electrons Backscattered Electrons
1
100 Energy of Electron (eV)
10,000
(Incident beam energy : 10,000eV)
Energy spectrum of the electrons emitted from a specimen
Electron Beam generated from Sample Primary beam
BSE Detector
High resolution Surface feature
BSE information SE information Composition information Crystal Orientation
S E Detector
less than 10nm Sample de pth of SE generating
sample depth of BSE ~ more than 10nm gene rating
Sample
Theory of Scanning Electron Microscope Tungsten Filament
FE Tip
750μm Electron Source Type of Emission Operating Vacum (Pa) Brightness (A/cm ・str) Source Size (μm) Energy Spred (eV) Life Time (h) 2
Tungsten Filament Thermonic 10-5 5x105 30 2.0 50
Field Emission Cold FE ∼10-8 108 0.01 0.2 2000
Comparison of electron sources
Theory of Scanning Electron Microscope Tungsten Filament
FE Tip ΔV= ∼0.2eV
ΔV= ∼2eV
Crossover of High Energy Electrons
Crossover of Low Energy Electrons
Crossover of High Energy Electrons
Energy Spread Effect of chromatic aberration
Crossover of Low Energy Electrons
Tungsten (W)
Op. T emp
2300deg. Operation
Lanthanum Hexaboride (LaB6)
Schottky FE
ColdColdCathode FE
1800deg. Operation
1800deg Operation
Ambient Temp
Energy spread
less than 2.0eV
less than 1.5eV
Less than 0.8eV
Less than 0.2eV
Brightness
5x105 A/cm2sr
5x106 A/cm2sr
5x10 8 A/cm2sr
2x109 A/cm 2sr
less than 100hr
500 – 1000 hr
1 year
More than 2 years
Tip change
Gun exchange
No Gun exchange
Tip activation
N eed continuously
Life time
Comparison Of Electron Sources
Need at time of use
Theory of Scanning Electron Microscope Primary beam
S-4300 SE Detector
Lens Specimen
1)Conventional type(Out-Lens)
S-4500 SE Detector (Upper)
S-5500 Primary beam SE Detector
Primary beam Lens
SE Detector (Lower)
Specimen
2)S nokel type
Lens
Specimen
3)In-lens type
Theory of Scanning Electron Microscope 20
Resolution (nm)
10
W filament SEM
Out lens FE-SEM
1.0
Snorkel lens FE-SEM 0.5 0.5
In-Lens FE-SEM 1.0
10
30
Acc.(kV)
Comparison of resolution
Reliable FE-Gan
1st anode FE tip
- Hitachi is manufacturing both FE Gun and FE Tip. - Baking is almost unne cessary afte r the installation.
2nd anode INNER baking OUTER baking
- Even if re quire d, it is very easy including inne r baking. - Heate d Obj. aperture can eliminate contamination for long life-time . - Hitachi’s FE tip long life has been well known as 3-7 years.
Adjustment of Accelerating voltage High Vacc(10kV∼30kV) 1)When thick metal layer is coated on the sample 2)When high resolution observation 3)When internal information of sample is required
Middle range Vacc(3kV∼10kV) 1)When high resolution and surface info rmation is required 2)When need good contrast on uncurved surface sample
Low Vacc(0.5kV∼3kV) 1)When surface feature observation 2)When need to avoid Charge-up effect and sample-damage
Beam spreading in the sample Vacc 1kV
Vacc 15kV
Magnified
20 nm
1μm 1μm
20 nm
a) Sample :Carbon Monte Carlo Simulation
Beam spreading in the sample Vacc 1kV
Vacc 15kV
Magnified
0.2μm 5 nm 5 nm
0.2μm
b) Sample :Gold
Monte Carlo Simulation
Theory of Scanning Electron Microscope Vacc:1.5kV
Chage-up Phenomena
Vacc:0.7kV
Eliminate Chage-up Phenomena Specimen : SiO2 on Photo Resist Line Pattern
Observation at lower accelerating voltages
Theory of Scanning Electron Microscope Vacc : 15kV
High Acclerating Voltage
Vacc : 1.0kV
Low Accelerating Voltage Specimen : Solar Battery
Comparison of high and low acclerating voltage
Electrical field effect to primary electron(E)
Primary elec tron Primary elec tron
Pr im ary e lect ron Pr im ary e lect ron
SE SEDetector Detector
ee
FFE E
Sample Sample
+1 0k V +1 0k V
Sample Sample
History of Upper SE Detector (+10kV has been applied to detector surface ) Instrument
Semi-inlens S EM
Inlens S EM
TEM
H V >75kV
Countermeasure No problem due to high HV
S-900
0.5 -30kV
Keep the detector away from axis
S-5000
0.5 -30kV
Insert shielding cylinder
S-5200
0.5 -30kV
S-4500
0.5 -30kV
S-4700
0.5 -30kV
S-4800
0.5 -30kV
ExB equipped
Patent No. : P3081393
What is ExB ? Current
I
Magnetic Field
B
Force
Primary electrons
E
SE Detector
B FB
e
FE +10kV
+V
FB Fleming’s left-hand rule
|FE|=|FB| Current(I) direction is opposite to electrons
OBJ Lens Sample
Patent No. : P3081393
What is ExB ?
Primary electrons
Magnetic Field B
E
SE Detector
B Force
FB Current
I
FB
e
FE +10kV
FE +V
e
Secondary Electrons
FB
Fleming’s left-hand rule
<Secondary electron case>
OBJ Lens Current(I) direction is opposite to electrons
Sample
Adjustment of Optics conditions d1(Beam spot size)= Gun
d0
d0 ・ ba11・ ba22
b1:Adjusted by Cond. lens
a1 Cond. Lens b1 Obj. Lens
a2 b2
Obj. lens aperture
d1 Sample
b2:Adjusted by WD
Relation between Cond. Lens and image Beam spot size can be adjusted by changing electric current of Condenser Lens.
Lens current
Small excitation
Large excitation
Long
Short
Resolution
Low
High
Beam current
Large
Small
S/N of image
Large
Small
b1
Relation between WD and image WD (Working Distance) : length between Obj. lens and sample surface WD
Long
b2
Long
Short
Resolution
Low
High
Lens excitation
small
large
Focus depth
Deep
Short
Shallow
Relation between WD and image Focus depth: Shallow Resolution : High
WD:5mm
Focus depth : Deep Resolution : Low
WD:20mm
Relation between Obj. lens aperture and image 1) Adjusting beam spreading angle 2) Adjusting beam current eradiating to sample Gun Shielded beam by Obj. lens aperture
Obj. Lens
Obj. Lens aperture W.D
Beam angle(α)
Sample
B2
Relation between Obj. lens aperture and image
Aperture Hole size
Large
Small
Resolution
Low
High
Beam current
Large
Small
S/N of image
Large
Small
Focus depth
Shallow
Deep
Theory of Scanning Electron Microscope Aperture Size : Small
Focus Depth → Deep
Aperture Size : Large
Focus Depth → Shallow
Specimen : Si on Photo Resist Pattern
Comparison of objective movable aperture hole size
Detecting Affection under SEM imaging Image detecting affection and countermeasures Affection
Charge-up effect
Contamination
Beam damage
Phenomenon
Countermeasures
*Extraordinary contrast *Drift of sample
*Metal coating *Observe at Low Vacc * Observe with BSE
*Less contrast
*Heating / Cooling *UV eradiation *Plasma cleaning
*Change of sample shape
*Metal coating * Observe at ultra Low Vacc. *Cooling Observation
Charge-up effect
What is Charge-up effect? Iin :electron flow entering sample = IP Iout :electron flow radiating from sample = IS E + IBS E + Iab Probe current (IP)
If the sample is conductive Iin = Iout となり Electric charge of sample is balanced
BSE flow (IBSE)
SE flow (ISE)
Sample
Absorbed electron flow (Iab)
What is Charge-up effect? Iin :electron flow entering sample = IP Iout :electron flow radiating from sample = IS E + IBS E + Iab Probe current (IP)
If the sample is non-conductive Iin ≠ Iout となり Charge-up effect
BSE flow (IBSE) Electric charge
SE Flow (ISE) ee- e- ee- e-
Sample
Absorbed electron flow (Iab)
Countermeasures against charge-up Metal coating Metal coating by vacuum elaboration system or Ion sputtering system Coating Material Need to select optimum material for SEM observation 1. Can coat at homogeneous distribution and fine particle 2. Good efficiency of SE generating 3. Stable against oxidization
Observation:Au、Au-Pd、Pt-Pd、Pt .... Analysis:C、Al....
Coating method Focus point in coating work Need to avoid affecting sample shape Artifact Changing sample size Contamination / Damage
Need to minimize
Coating layer
Sample Homogeneous distribution
Not Homogeneous distribution
Coating method Artifact by coating treatment
200nm
Non coating
Pt coating
Vacc :2kV
Vacc:3kV
Coating method Less than 3nm coating layer is ideal to avoid artifact
100nm
Less than 3nm
How about odd-shaped sample?
Coating method a) Coating from only overhead
Coating layer
b) Coating from several direction
Coating direction
Sample
*Coating layer become thick when you want to avoid charge-up *Surface feature is not accurate
* Surface feature is accurate
Imaging technique What is “ Just focus” focus”?
Condensed electron beam to finest beam spot is eradiated at sample surface
Spot shape should be “perfect circle” circle ”
Theory of Scanning Electron Microscope Before correction
Y Beam Diameter X
Electron Source
Electron Beam Objective Lens
After correction
Stigmator
Y
Beam Diameter X
Electron Source
Electron Beam Stigmator
Objective Lens
Astigmatism correction method
Theory of Scanning Electron Microscope Under focus
Just focus
Over focus
Before correction Just focus
After correction
Specimen:Trachea of rat
Astigmatism correction method
Ideal adjustment of focus and stigma stigma
X
focus
Y
Beam Beamspot spotshape shape Ideal beam spot shape
Present beam spot shape
coarse
fine
Ideal adjustment of focus and stigma stigma
X
Beam Beamspot spotshape shape
focus
Y
coarse
fine
Ideal adjustment of focus and stigma stigma
X
Beam Beamspot spotshape shape
focus
Y
coarse
fine
Ideal adjustment of focus and stigma Adjust knob to center where image is not drifting
stigma
X
Beam Beamspot spotshape shape
focus
Y
coarse
fine
Ideal adjustment of focus and stigma stigma
X
Beam Beamspot spotshape shape
focus
Y
coarse
fine
Ideal adjustment of focus and stigma stigma
X
Beam Beamspot spotshape shape
focus
Y
coarse
fine
Summary: Imaging technique for high resolution Gun
d1= d0 ・ ba11・ ba22 ・ ba33
d0 a1 1st Cond. lens b1
How to minimize d1?
a2 2nd Cond. lens b2
Shorten b1: Increasing excitation of Cond. lens
a3 b3
d1 Sample
Obj. Lens (Focus knob)
Shorten b3: Shorten WD
Summary: Imaging technique for high resolution
Cond. Lens:1notch WD:12mm
Cond. Lens :8 notch W D:2.5mm
Sample:ITO layer Vacc :3.0kV Mag. : x 100,000